หัววัดค่า pH(pH electrode) pH electrode (5-metre cable and plastic BNC connector)พร้อม Probe

คุณสมบัติสินค้า:

SKU : 21538-01306

หัววัดค่า pH พร้อม Probe (update 30/08/2024)

หมวดหมู่ : Sparepart Astral Control Dosing Probe

แบรนด์ : ASTRALPOOL

Share

pH Probe Sensor หัววัดค่า pH  พร้อมหัวยึด(Probe)

Accessories for pumps in the Class range
Includes Probe, Probe Holder, Support Collar and 1 Buffer Solution

   -pH electrode (5-metre cable and plastic BNC connector) with Probe

 


 This pH electrode is a hand crafted, precision analytical device. Carefully follow the directions in this instruction sheet to obtain the best performance and electrode life.

Required Equipment and Solutions

   ·  pH meter (a temperature compensator is recommended for increased pH measurement accuracy)

·     pH Buffer 4.01
·      pH Buffer 7.00
·      Reference filling solution (see list for particular pH electrode type)
·      Wash bottle filled with distiled or de-ionized water
·      Laboratory magnetic stirrer and magnetic stir bars
·      Lab wipes
·      Clean beakers

Preparation of the Electrode for Initial Use

1.    Remove the protective bottle or cover from the electrode and thoroughly rinse the electrode with distilled water. Wipe carefully with a clean lab wipe.

2.    During shipment, air bubbles may have migrated into the electrode sensing bulb. Hold the electrode up to the light and inspect the sensing bulb for air bubbles. If air is seen, carefully shake the electrode downward (like a thermometer) to dispel the air bubble from the sensing bulb at the tip of the electrode.

3.    For refillable models, uncover the filling port to expose the electrode reference chamber fill hole (for sealed, gel filled electrodes, disregard this operation).Fill the reference chamber with the appropriate pll reference filling  solution. Electrodes that have been filled with the incorrect filling solution are not covered under warranty.

Reference Filling Solution Selection
   ·  For pH combination electrodes with Calomel and Double Junction  Ag/AgCI reference half cells, use 4 M KCI reference filling solution.
   ·   For pH combination electrodes with Single Junction Ag/AgCI reference half cells, use 4 M KCI saturated with AgCI reference filling solution.

Standardizing the Electrode
   1.    Connect your pH electrode to the input connector on the pH meter. Ensure that the electrode connection is secure.
   2.    Place the electrode into a beaker containing pH 7.00 buffer and a stir bar. Stir at a moderate speed.
   3.    When the reading is stable, adjust the meter to read the value of the pH7.00 buffer at the temperature in the laboratory (see the meter manufacturer's instructions for details regarding temperature compensation)
   4.    Remove the electrode from the buffer. Rinse with distilled water and blot with a lab wipe.
   5.    Place the electrode in a beaker containing pH 4.01 buffer and a stir bar. Stir as before. When the reading is stable, adjust the meter to read the value of the 4.01 buffer at the temperature in the laboratory.


Reading a Sample with the Electrode
   1.    Rinse the electrode with distilled water and blot with a lab wipe. Place the electrode in a beaker containing the sample and a stir bar. The  sample should be at the same temperature as the buffers used to  standardize the electrode. Stir as before. Record the pH when the reading is stable.
   2.    Remove the electrode from the sample, rinse the electrode with distilled water over the "waste" beaker. Blot the electrode dry with a  lab wipe. The electrode is now ready to read the pll of other samples.

Storing the Electrode

Sort Term
   Between measurements, store the pH electrode in a beaker containing pH 4.01 buffer.

Long Term
   When storing for longer periods, store the pHI electrode in the storage bottle or the protective boot which came with the electrode. Ensure that the foam in the storage bottle or the cotton ball in the protective boot is totally wetted with DI water to keep a moist environment around the pH bulb and junction.
Maintain the moist environment in the storage bottle or in the protective boot during storage.

If the pH electrode is equipped with a fill hole cover, slip the cover over the fill hole.

Electrode Cleaning
   Do not use strong solvents (e.g. acetone, carbon tetrachloride, etc.) to clean the pH electrode. Be sure to recalibrate the electrode after cleaning.
   1.  If the electrode has become coated with oil or grease, carefully wash the electrode under warm tap water using dish-washing detergent.  Rinse thoroughly with fresh tap water followed by a rinse with distilled water. Soak the electrode in pH electrode storage solution for 30 minutes after this cleaning procedure. Recalibrate the electrode before use.
   2.  If the electrode has been exposed to protein or similar materials, soak in acidic pepsin, The part number CS 0003 for 5 minutes. Rinse thor-oughly with distilled water. Recalibrate before use.
   3.  If the previous cleaning procedures fail to restore response, soak the electrode in 0.1 N HCI for 30 minutes. Rinse thoroughly with distilled water. Recalibrate before use.
   4.  If electrode response is not restored still, replace the electrode.